INSTRUCTIONS FOR USE
4. To adjust height, loosen the rail clamp knobs. Grasp the table top adjacent to the
rail clamps and raise or lower the table to the desired height. Securely tighten the
rail clamp knobs.
5. Set the table at the surgeons working height, loosen Leg Lock Handle at the ball
socket near end of table. Swing the leg down and slide ball socket to position of
least interference with surgeon, C-arm, etc. Loosen Leg Extension Knob and
adjust leg length so the foot is securely against the floor. Firmly lock both handles
to stabilize table end.
Option for 30410-C
and 30450-C
Device controls and indicators:
Controls and indicators of this device are described within the Setup instructions.
Storage, Handling and Removal Instructions:
3.4.1
Storage and Handling:
The product should be stored in a clean and safe environment to prevent product
damage. See storage Specifications under Product Specification section.
3.4.2
Removal Instruction:
1. Loosen the knob and remove the table from the clamp.
2. Remove the rail clamps form the surgical table.
Troubleshooting Guide:
This device does not have a troubleshooting guide. For technical support user of the
device shall first contact Hill-Rom Technical Support.
Document Number: D-720113
Version: C
The Optional Leg A-30450-LEG is required when
the Allen® Carbon Lights™ Minor Procedures
Table and Rectangular Table are used with the
Schlein Hand Positioner (A-20750).
Page 13
Issue Date: 26 MAR 2020
Ref Blank Template: 80025117 Ver. F