² Do not use this battery on other devices;
² Do not use this battery below -10 ℃ or above 40 ℃ ;
² Dispose of the battery, the local law should be followed.
% In order to maintain battery supply time and prolong battery lifetime, please charge the battery every
one or two months if don't use battery for a long time. And do charge battery at least 12-15 hours
every time. Before connect to AC, do start monitor with battery's power supply, until battery power
is used up and monitor turn off automatically, then connect monitor to AC and have it charged for
12-15 hours continuously. The speed of charge will be the same no matter whether the monitor is
working or not. The reason why discharge the battery before charge is to avoid the decrease of
capacity caused by battery's memory effect. If the monitor won't be used for a long time, do have it
charged fully before conservation.
% Using a monitor powered solely by an internal battery power which has short charge power will
cause the monitor turn off automatically when the battery is depleted.
% If battery is damaged, please replace with same type and specification battery marked by "CCC" or
"CE" in time, or contact the company directly.

10.1.4 Service

If the monitor has functional malfunction or is not working, please contact the local dealer or our company, and
we are to offer the best solution as soon as possible for your satisfaction. Only qualified service engineer
specified by the manufacture can perform the service. Users are not permitted to repair it by themselves.

10.2 Cleaning and Disinfection

n Kept the monitor from dust.
n It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only non-corrosive cleanser
such as clear water is permitted.
n Use the cloth with alcohol to wipe the surface of the monitor and transducers, and dry it with dry and
clean cloth or simply air-dry.
n The monitor can be sterilized and disinfected, please clean it first.
M Switch off the monitor and disconnect the power cable before cleaning.
M Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.
M Clean the exterior of the connector only.
% Dilute the cleanser.
% Do not let any liquid flow into the shell or any parts of the monitor.
% Do not let the cleanser and disinfectant stay on its surface.
% Do not perform high pressure sterilization to the monitor.
% Do not put any parts of the monitor or its accessories in the liquid.
% If the monitor is accidentally wetted it should be thoroughly dried before use. The rear cover can be
removed by qualified service technician to verify absence of water.