Measurement of the Light Intensity
The charging unit of the Translux 2Wave is equipped with an integrated light intensity meter ( 8 ).
CAUTION
Before each treatment always check the light intensity by using the light intensity meter
integrated in the Translux 2Wave charging unit.
• Without applying pressure, place the light guide flat and directly on the testing area ( 8 ).
Activate the unit by pressing the " STAND-BY "
The light intensity measuring is immediately indicated by the 2-colour LED (Symbol
of the testing area.
GREEN LED : The light intensity is sufficient for the treatment.
YELLOW LED : The light intensity is insufficient and patient/s must not be treated.
CAUTION
If the light intensity is insufficient, please carry out the following controls and measures:
– The light guide has not been inserted into the hand piece (see paragraph 5.4 Inserting the fibre optic
into the hand piece).
– Check the light guide for contamination or defects.
– Clean the light guide of any contamination (see paragraph 8 Cleaning, disinfection and sterilisation)
or replace the damaged light guide with a new one.
Use original light guides only. Kulzer GmbH will provide only original light guides supplied by Mectron S.p.A.
WARNING
If the measures described above do not result in any improvement, do not use the device for any treatment.
Switch off the unit (disconnect the charging unit from the electrical power source) and ensure that
unauthorised unit re-activation is not possible. Contact a Kulzer and/or authorized Kulzer service partners
only (see paragraph 14 Service).
6.4
Sleep Mode
The hand piece is equipped with a " Sleep Mode " to minimise the unit's energy consumption.
Outside the charging unit, all internal functions of the hand piece are automatically turned off
(" Sleep Mode ") if it is not used for approximately 5 minutes.
In " Sleep Mode " the hand piece display is in the condition shown in Fig. e :
To terminate the " Sleep Mode " press the " STAND-BY " button
The hand piece is ready for operation.
NOTE
Terminated the " Sleep Mode" , the hand piece and the display return to the latest setting used.
6.5
Forced Sleep Mode
Outside the charging unit it is possible to switch the hand piece to the " Sleep Mode " at any time.
• To activate the " Forced Sleep Mode " press and keep pressed for 5 seconds the " Time / Program " button
The hand piece and the display switch to the same conditions described in the above paragraph 6.4 (Sleep Mode).
• To terminate the " Sleep Mode " press the " STAND-BY " button
The hand piece is ready for operation.
NOTE – Memory Function
By terminating the " Sleep Mode ", the hand piece and the display return to the latest setting used.