Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
1. Press the ■ button to enter setup.
NOTE: The screen for the first calibration cylinder displays.
2. Press
○ the ▼ or ▲ button to change the value.
○ the ■ button to confirm the setup.
NOTE: With this confirmation the device automatically moves to the next cylinder setting.
3. Repeat the sequence for changing the required settings for all necessary gas values.
NOTE: After the last setting is performed, the device returns to the Calibration Options menu.
Setting Cal Due Options
1. Press the ■ button to enter setup.
2. Press the ▼ or ▲ button to enable/disable this option.
3. Press the ■ button to confirm.
4. After confirmation the device prompts the user to enter the number of days for the reminder.
5. Change number of days by pressing the ▼ or ▲button.
6. Press the ■ button to go to the next menu.
Setting Last Cal Date
1. Press the ■ button to enable/disable this option.
2. Press the ▼ button to go to the next page.
3. Press the ▲ button to go to the previous page.
Setting Calibration Password
1. Press the ■ button to enable/disable this option.
2. Press the ▼ button to go to the next page.
3. Press the ▲ button to go to the previous page.
Back To Main Menu
1. Press the ■ button to go to Device Setup Menu
NOTE: The Cal Options screen displays
2. Press the ▼ button to go to the next (Alarm options) or the ▲ button to exit the Setup menu.
4.5.2
Alarm Setup
Alarm Options
US
ALTAIR 5X / ALTAIR 5X IR
4 Operation
25