6 Technical Specifications
6
Technical Specifications
Weight
Weight (with IR Sensor)
Dimensions (cm)
Dimensions (cm) (with IR Sensor)
Alarms
Volume of audible alarm
Displays
Battery types
Charging time
Normal Temperature range
Extended Temperature range
Short Term Operations (15 minutes)
temperature range
Humidity range
Atmospheric pressure range
Ingress protection
Measuring methods
Warranty
CIO
2
Cl
2
CO
CO - HC
Combustible
H
S
2
H
S - LC
2
HCN
42
0.45 kg (1 lb.) - device with battery and clip (ALTAIR 5X unit)
0.52 kg (1.15 lb.)
17 x 8.87 x 4.55 (6.69" H x 3.49" W x 1.79" D) pumped, without belt clip (ALTAIR 5X unit)
17 x 8.94 x 4.88 (6.68" H x 3.52" W x 1.92" D)
LEDs, audible alarm, vibrating alarm
95 dB typical
Monochrome / Color
Rechargeable Li ION battery
Replaceable AA alkaline (ALTAIR 5X only)
For battery types see the Altair 5X XIR User Instruction Addendum (P/N 10223097)
≤ 6 hours
The maximum safe area charging voltage
Um = 6.7 Volts DC
-10 °C to 40 °C (14 °F to 104 °F)
-20 °C to 50 °C (-4 °F to 122 °F) Monochrome display
-10 °C to 50 °C (14 °F to 122 °F) color display
-20 °C to 40 °C (-4 °F to 104 °F) for devices with ClO
-40 °C to 50 °C (-40 °F to 122 °F) for devices without PID sensors
15 - 90 % relative humidity, non-condensing,
5 - 95 % RH intermittent
80 kPa to 120 kPa (11.6 to 17.4 PSIA)
IP 65
Combustible gases - Catalytic or Infrared sensor
Oxygen and Toxic gases - Electrochemical or Infrared sensor
See
2.4 Warranty
Measuring Range
0-1.00 ppm
0-10 ppm
0-2000 ppm
0-10,000 ppm
0-100 % LEL
0-5.00 % CH
4
0-200 ppm
0-30 ppm
ALTAIR 5X / ALTAIR 5X IR
sensors
2
NH
3
NO
NO
2
NO
(XCell)
2
O
2
PH
3
PID
SO
2
0-100 ppm
0-200 ppm
0-20.0 ppm
0-50.0 ppm
0-30 % Vol.
0-5.00 ppm
0-2000 ppm
0-20.0 ppm
US