4. Use both the paper-like filter and the fibrous dust filter (the thicker disk) as supplied in Maintenance Kit if the device is
NOT configured to use a reactive toxic gas sensor (does not have a Cl
NOTE: Use ONLY the paper filter supplied in the Reactive Gas Maintenance Kit if the device IS configured to use a
reactive toxic gas sensor (Cl
5. Place the new paper-like filter into the recess in the back of the device. If it is to be used, place the fibrous dust filter
into the clear filter cover.
WARNING!
Use of the fibrous dust filter or the incorrect paper filter for the measurement of reactive gases could cause erroneous
readings.
Failure to follow this warning can result in serious personal injury or death.
6. Replace the O-ring in the recess.
7. Re-install the clear filter cover on the back of the device.
5.6
Cleaning the Device Exterior
Clean the exterior of the device regularly using only a damp cloth. Do not use clea-ning agents as many contain silicones
which will damage the combustible sensor.
5.7
Storage
When not in use, store the device in a safe, dry place between 18 °C (65 °F) and 30 °C (86 °F). After storage, always
recheck device calibration before use. If not to be used in 30 days remove battery pack or connect it to a charger.
5.8
Shipment
Pack the device in its original shipping container with suitable padding. If the original container is unavailable, an equivalent
container may be substituted.