Chauvin Arnoux C.A 6549 Manual De Instrucciones página 71

Megaóhmetro
Ocultar thumbs Ver también para C.A 6549:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

§ DAR (s : s)
§ PI (m : m)
§
Set Step Function 1, 2, 3
Step Function 1
Step Function 2
Step Function 3
Note : the minimum sample time is related to the total duration of the test (Total Run Time). It is equal to
: Sample Time (seconds) = (h+1)*5 where h= total run time in hours.
§ Temperature Unit
1
st
and 2
nd
times for the DAR calculation
Default value
30 : 60 (s:s)
st
nd
1
and 2
times for the PI calculation
Default value
01 : 10 (m:m)
for each predefined step function mode, definition of the
various voltages, of the duration of each step, and of the
duration for the recableing of samples.
Default value
Voltage
Duration (h:m)
step 1
50V
step 2
100V
step 3
150V
step 4
200V
step 5
250V
sample time
00 : 01 (m:s)
step 1
100V
step 2
300V
step 3
500V
step 4
7000V
step 5
900V
sample time
00 : 01 (m:s)
step 1
1000V
step 2
2000V
step 3
3000V
step 4
4000V
step 5
5000V
sample time
00 : 01 (m:s)
selection of temperature unit
Default value
°C
71
Range
10... 90 : 15... 180 (s:s)
5-second steps
Range
0,5... 30
(0.5-then 1-mn steps)
1... 90 (
0.5, then 1-, then 5-mn steps)
Range
Voltage
00 : 01
40V à 5100V
(in 10-V, then 100-V
00 : 01
steps)
00 : 01
00 : 01
00 : 01
00 : 01
40V à 5100V
(in 10-V, then 100-V
00 : 01
steps)
00 : 01
00 : 01
00 : 01
00 : 01
40V à 5100V
(in 10-V, then 100-V
00 : 01
steps)
00 : 01
00 : 01
00 : 01
Range
°C or °F
Duration (h:m)
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
see note
(00... 59 : 0...59)
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
see note
(00... 59 : 0... 59)
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
00... 09 : 01... 59
see note
(00... 59 : 0...59)

Publicidad

Tabla de contenido
loading

Tabla de contenido