12
Troubleshooting
12.1
Command 9: Read device variables with status
Byte
Bit
Format
0
0
Bit
1
2
3
4
5
4-7
Float
8
6 and 7
Bits
Instructions for use
|
Dräger Polytron
Extended device status
Description
Maintenance required
WARNING_UNDER-
RANGE_ANEG
ERROR_UNDERRANGE_ANEG
Alarm device variable
All alarms and warnings
Critical current output
WARNING_BATTERY_LOW
ERROR_BATTERY_CRITICAL-
LY_LOW
STA_ERR_ADC_REF
STA_ERR_ADC_EXTREF
Error
All errors
Outside the specification
TEMPERATURE_TOO_HIGH
TEMPERATURE_TOO_LOW
Functional test
MAINTENANCE_ENABLED
SIMULATION_ACTIVE
Device variable value
Device variable status
Meaning of "Quality" (bit 6 and 7):
3: Good - device variable value with
normal accuracy.
2: Poor - device variable value with
low accuracy.
1: Manual/Fixed - fixed device vari-
able value.
0: Bad - there may be a device vari-
able value, but the device variable
value may not be used.
®
6700 IR WL, Dräger Polytron
Cause
Under range.
Fallen below negative threshold.
Battery charge low.
Battery charge critical.
Operating voltage too low.
Operating voltage too low.
Device is operated outside the
specified temperature range (upper
limit).
Device is operated outside the
specified temperature range (lower
limit).
Maintenance mode active.
Device is in simulation mode.
®
®
Repeater WirelessHART
|
en
Troubleshooting
Remedy
Performing zero adjustment.
Performing zero adjustment.
Change battery pack soon.
Change battery pack immediately.
Check voltage.
Check voltage.
In case of high temperatures due to
solar radiation, use weather protec-
tion set (accessory).
In case of low temperatures, pay
attention to operating temperature.
Exit maintenance mode.
Exit simulation mode via PolySoft.
51