Byte
Bit
Description
16
0
OUT_OF_SERVICE
1
INSTALLATION_CALIBRATION
2
FAULTS_SENSOR
3
FAULTS_ELECTRONICS
4
M: Maintenance required
5
O: Out of specification
6
C: Function Check
7
F: Failure
17
4
PROGRAM_FLOW_ERROR
19
2
DEV_CONFIG_CHECK
Instructions for use
|
Dräger Polytron
Cause
Start sequence active.
Start sequence failed.
Adjustment or bump test active.
Sensor defective.
Incorrect power supply setting (battery or wired).
Serious device error.
–
–
–
–
Error in electronics.
Error in electronics.
®
6700 IR WL, Dräger Polytron
Remedy
Wait for start sequence.
A
Wait up to 15 minutes for the start sequence to
complete.
B
Restart the device.
C
Check the condition of the battery. Test with new
battery pack.
D
Contact DrägerService.
E
(if applicable) complete or repeat gas library
update.
Finish adjustment or bump test.
A
Restart the device.
B
Check the condition of the battery.
C
Contact DrägerService.
A
Check the setting of the power supply.
B
Restart the device.
C
Reset device to factory settings.
D
Contact DrägerService.
–
–
–
–
A
Restart the device.
B
Reset device to factory settings.
A
Reset device to factory settings.
B
Contact DrägerService.
®
®
Repeater WirelessHART
|
en
Troubleshooting
53