2) Enter a value from O to 9 on the numeric keypad.
3) Press f-- and
to move the cursor, and press 1'
4) Press ENTER to turn ON/OFF.
3.2.2.
Static CV
In constant voltage mode, the load keeps the device under test at the set voltage regardless of whether
the input current is changed.
Operations: same as above
3.2.3.
Static CW
In constant voltage mode, the load keeps the device under test at the set power regardless of whether the
input voltage and current are changed. Operations: same as above
3.2.4.
Static CR
In constant resistance mode, the load keeps the device under test at the set resistance regardless of
whether the input voltage and current are changed. Operations: same as above
3.2.5.
SHORT function
In SHORT made, the load is output at the maximum current. Operations:
Long press CC to displays MODE: SHORT on the interface and press CC, CV, CW to exit.
3.2.6.
Call static storage
The load can stare and call 100 groups of static set values.
1) Storage operation
(1) Press CAL/SAVE to switch the state to SAVE.
(2) Enter a number key to index a line in the list, press TAB to select it, and press ENTER to enter the edit
made. Edit to display a red background, and press ← and → to select.
(3) Edit MODE. Press CC, CV, CW, or CR on the keyboard to modify.
(4) Edit data. Press 0 to 9 and ESC on the keyboard to modify.
(5) After modification, press ENTER, and long press ENTER again to save the data.
2) Call operation
(1) Press CALL/SAVE to switch to CALL.
(2) Enter a number key to index a line or press TAB to switch the list, select it with the knob, and press
ENTER.
3) M1 to M6 quick call
and knob to adjust the corresponding value.