Turn the knob to adjust the brightness, long press ENTER or numeric 8 to save, and press ESC to exit.
3.10.10.
0 calibration function
If the device has not been calibrated for a long time or the temperature is high or low, the data will not
return to zero, and the 0-calibration function can be performed.
After entering home page, close the load, unplug the external test line, and long press numeric 0 for over 3
seconds for the 0 calibration of voltage and current. And the 0-calibration function will disappear when the
load restarts, and 0-calibrate again when needed.
3.11.
U Flash Disk Import and Export Function
Save real-time voltage and current data and export data list through U flash disk. As shown in Fig. 11.1, the
USB[0] tree on the left shows that the file list is the acceptable format data loaded in the U flash disk. And
the upper part on the right is the option of importing the load from the U flash disk, and the Iower
part is the data option of exporting the load to the U flash disk.
Press TAB to select the control. Press ← → ↑ ↓ to move the focus left or right.
Press F4 (Load CSV) to import the file from the USB flash disk, and save CSV (FS) to export the CSV file to
the USB flash disk. And the function keys are shown in Fig. 11.2 below:
Start SaveData(Fl)
Stop SaveData(F2)
USB
Disconnect(F3)
LoadCSV{F4)
SaveCSV{FS)
HomePage(F6)
3.11.1.
Import and save LIST
Take the table that data of BAT Group 1 is exported as an example.
Fig. 10.3 Backlight
Fig. 11.1 Storage Interface of USB Flash Disk
Create a CSV file with date as its na me
Stop writing files
Disconnect USB
Import files from USB flash disk
Export CSV file
Home page