3.6.
VI Curve Mode
A maximum of 7 groups of VI test parameters can be set according to the set maximum current, minimum
current, and step value.
Fig. 6.1 VI Current Parameters
3.6.1.
VI test setup
Operation instructions: it takes the settings stored in Group 1, the start current of 0.2A, the end current of
SA, the step current of 0.2A, and the step duration of ls as an example.
1) Press (F3) Battery on the main interface to enter VI measurement.
2) Press (F3) Curve/CALL, and press TAB to select CALL GROUP 1
3) Press (F4) Curve/SAVE to display SAVE on the table.
4) Press TAB to switch to the table focus and turn the knob to select the row position to be modified.
5) Press ENTER to select the red background.
6) Press ENTER to select the red background.
7) After editing as 0.2, press ENTER. Then 0.2000 will be displayed on the screen, and the background will
be switched to blue.
8) Repeat the above steps to set 5A, 2A, and 1.000s respectively.
9) Long press ENTER to save the edited data.
10) Press (F3) Curve/CALL.
11) Press ENTER to call.
12) Turn ON/OFF.
The operation effect is shown in Fig. 6.2