STAMOS S-LS-118 Manual De Instrucciones página 35

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 123
Fig. 6.2 VI Running Interface
3.7.
OCP Mode
When the voltage reaches the VON value, the current output will be delayed for a period of time, and the
step value will be decreased once every other period of time until the cut-off current or the voltage is
higher than the OCP set voltage. If the stopped voltage is higher than the OCP voltage and the current is
between the set maximum and minimum value, it is PASS, otherwise it is FAULT.
3.7.1.
Setting function of OCP test
Note: 7 groups of OCP test parameters can be set at most.
Operation instructions: it takes the setting stored in Group 1, VON voltage of 10V, VON voltage delay of 5S,
and current range of 3A; start current of 2A, each decrease of 0.1A, and each decreasing duration of 1s;
end current of 1A, OCP voltage of 8V, maximum current of 1.9A, and minimum current of 1,1A as an
example.
Fig. 7.1 OCP
1) Press (F2) OCP/OPP on home page.
2) Press (F1) OCP/CALL and TAB to select CALL GROUP 1.
3) Press (F2) OCP/SAVE to display SAVE on the table.
4) Press TAB to switch the focus and turn the knob to select the first row.
5) Press ENTER to select red background. Enter numeric O to 9, and ESC.
6) After editing VON as 10V, press ENTER. Then 10.000 will be displayed on the screen, and the
background will be switched to blue.
7) Repeat the above steps to set 5S, 3A, 2A, 0.1A, 1S, 1A, 8V, 1.9A, and 1.1A respectively.