Descargar Imprimir esta página

Stryker 8000-060-006 Instrucciones De Uso página 16

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 73
6.
Disassembly
1. Unscrew and remove the knurled screw (Figure 1, A).
2. Detach the tracker from the instrument clamp.
7.
Product Lifetime
The instrument is validated for ten reprocessing cycles. Thus, the instrument contains a
usage counter that disables the instrument after running the tenth usage cycle in surgery.
The usage counter is updated as soon as the instrument is connected during surgery and
shown in the software.
To maintain accurate use count, connect the instrument to the electromagnetic port at
least once during surgery. This also applies to replacement instruments if they have been
taken from sterile packaging.
14 | Product Lifetime
EN

Publicidad

loading