LED
LED
Battery char
on
on
ger connected
power
bat
to product
sup
tery
ply
char
ger
Yes
Beep signal
4 x short at intervals
of approx. 20 sec.
(continuously)
13.2.3 Status signals
Battery charger connected
LED on
LED on
power
battery
supply
charger
Battery charger disconnected
Beep
Vibration
signal
signal
1 x short
1 x short
3 x short
Error
Distance between battery
charger and receiver of the
charging unit too big
No connection between bat
tery charger and power sup
ply
Defective battery charger
Error
Charging the battery outside the
allowable temperature range
Event
When connecting the power supply to the outlet, the LED on the battery
charger lights up briefly
Power supply and battery charger operational
Event
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery
charger. If the beep signal sounds again, you should visit your prosthet
ist soon. If necessary, the prosthetist will forward the product to an
authorised Ottobock Service Centre.
The product can be used without restrictions. However, vibration signals
may not be generated.
Resolution
The distance between the
battery charger and the
receiver on the product must
not exceed 1 mm / 0.04 inch
Check whether the charging
cable plug is fully engaged
on the battery charger.
The charger and the power
supply must be inspected by
a prosthetist.
Resolution
Check whether the specified ambi
ent conditions for charging the bat
tery are met (see Page 60).
65