IND780 Installation Manual
Sealing the Enclosure
When the IND780 terminal is used in a metrologically "approved" application, it
must be protected from tampering by use of seals. An optional sealing kit is
available from METTLER TOLEDO that contains all the required hardware (Part
number 64056538). Note that when the terminal is sealed non-metrological
components cannot be serviced without breaking the seal.
Figure 2-83 and Figure 2-84 show the components of the sealing kit and how to
use them, respectively.
Figure 2-83: Components of the Sealing Kit: Wire and Seal (left) and Screw (right)
Wire Through Screw (left), Wire Through Seal (center), Seal Closed (right)
2-48
Figure 2-84: Use of the Sealing Kit: