Harsh Enclosure Sealing - Mettler Toledo IND780 Manual De Instalación

Ocultar thumbs Ver también para IND780:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

IND780 Installation Manual
4. Finally, the back panel of the enclosure is sealed. Three sealing screws are
indicated in Figure 2-88 – two secure the panel in place on its stand-offs and
one secures one end of the upper card guide.
Screw securing
upper card guide
5. With the screws installed, thread the sealing wire through each of them, then
thread its ends through the plastic seal.
6. Remove most of the slack from the wire, and snap the seal shut.

Harsh Enclosure Sealing

For external sealing of the harsh enclosure, refer to Figure 2-89 and follow these
steps:
1. Ensure that the appropriate approval region has been selected in setup under
Scale, Type, Approval and that the Metrology security switch S-1 is in the "on"
position.
2. Thread the end of the wire seal through the hole in the bottom edge of the
IND780 front panel, and through the hole in the center clip of the IND780 front
panel.
3. Remove the slack in the wire seal and snap the front panel down to the
enclosure so that it snaps in place in all four corners.
4. Remove any remaining slack in the wire cable.
5. Thread the end of the wire cable through the plastic seal and snap the seal
shut. Note that, to show the relationship between the components, Figure 2-89
shows the cover open, the wire slack, and the seal not yet shut.
2-50
Figure 2-88: 3-Point Sealing, Panel Enclosure Rear Cover
Figure 2-89: External Sealing of the Harsh Enclosure
Rear panel
mounting screws
with sealing wire
Plastic seal

Publicidad

Tabla de contenido
loading