Mettler Toledo IND780 Manual De Instalación página 57

Ocultar thumbs Ver también para IND780:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

rear of the harsh enclosure. The other end of the flat braided cable is
terminated at the ground rod.
Metal clamp
Figure 2-44: PDX Home Run Outer Braid Termination, Harsh Enclosure
4. To terminate the home run cable's outer and inner braid shield for an IND780
panel mount terminal, when using a cable gland with the external panel
housing, follow the procedure described in step 3, above. Alternatively, when
using a conduit hub entry:
a. Slide back the outer braid and fold the trimmed inner braid and foil shield
back evenly over the edge of the cable insulation sheath, as shown in
Figure 2-45.
Inner drain wire
wrapped around
power ground wire
Foil shield
Outer braid
Figure 2-45: PDX Home Run Cable, Panel Mount Enclosure
b. Slide forward the outer braid to cover over the inner braid and foil. Pull the
outer drain wire back over the outer braid and trim it to approximately 25
mm (1 inch). Place the flat braided ground cable around the home run
IND780 Installation Manual
Flat braided cable
Outer drain wire
External grounding
point
Inner braid
Outer drain
wire
2-29

Publicidad

Tabla de contenido
loading