Modifying And Saving Programs - Boston Scientific Vercise Manual De Uso

Ocultar thumbs Ver también para Vercise:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Physician Manual
P
2. Press
or to move until the program you want is highlighted.
3. Press ▲ to activate the program.
Once you have selected and activated a program, the Remote Control will return to
the Level screen.

Modifying and Saving Programs

P
1. Press
as needed from the Level screen to cycle to the desired program.
Note: The most recently retrieved or saved program will be underlined(_).
Empty program slots are denoted by a blank box with no program number [].
If a patient tries to activate an empty program, nothing will happen.
2. Press ▲ to activate the selected program.
3. After the Remote Control times out to the Level screen, you may change the
amplitude of all areas in the program.
4. To save/store changes, select the program again and press ▼. You will be
required to confirm the adjustment first.
P
5. Press
to confirm, or press the button to cancel the operation.
Note: To save the modifications as a new program, simply select an empty []
program slot and press ▼ instead of overwriting the existing program.
Physician Manual
91098825-07 Rev A 52 of 333

Publicidad

Tabla de contenido
loading

Tabla de contenido