Homecare lock check (REF 66801496 only)
1.
Press SELECT button to start device.
2.
Simultaneously press and hold KEYPAD LOCK and
AUDIO PAUSE button for 5 seconds.
3.
Verify indicator light above KEYPAD LOCK button
illuminates solid blue, screen displays "Keypad
Locked", "Continuous" on the upper line and the
set-point followed by "***" on lower line.
4.
Press SELECT button to stop device. Verify device
does not stop and screen displays "Keypad Locked".
5.
Press UP and DOWN buttons in attempt to change
therapy. Verify device does not change and screen
displays "Keypad Locked".
6.
Press and hold POWER for 2 seconds to turn device
off.
7.
Press and hold POWER for 2 seconds to turn device
on. Verify device turns on, starts operating, and
displays therapy set-point followed by "***" on lower
line.
8.
Simultaneously press and hold the KEYPAD
LOCK and AUDIO PAUSE for 5 seconds to remove
homecare lock. Verify indicator light above KEYPAD
LOCK is no longer illuminated and screen displays
"Keypad Active".
9.
Press SELECT to stop device. Press and hold POWER
for 2 seconds to turn device off.
Battery function check
1.
Disconnect power supply from device and power
cord from AC mains power.
2.
Press and hold POWER button for 2 seconds to
turn device on. Verify during POWER button press,
status/alarm indicator illuminates solid yellow, and
all indicator lights illuminate.
3.
Verify display shows "Welcome, Starting", current
software version (e.g. V 1.06), and changes to show
"Standby" and therapy set-point.
4.
With device still set to 80mmHg, press SELECT
button. Verify device starts operating, status/
alarm indicator illuminates solid green, and screen
displays "Continuous" and therapy set-point.
5.
Press SELECT button to stop device.

Alarm indicator check

Recommended personnel: Alarm indicator checks
should be performed by a trained technician.
Equipment required: Stopwatch, RENASYS™ GO canister.
Complete blockage/canister over-capacity alarm
and audio pause button check
Note: Alarms can be tested on either battery or AC
mains power. If testing on internal battery power,
ensure internal battery charge is sufficient. See the
RENASYS GO user manual for details.
1.
Disconnect quick click adaptor from canister quick
click connector and insert tethered cap into canister
tubing.
2.
Reset stopwatch to zero.
3.
Adjust therapy set-point to 100mmHg using UP and
DOWN buttons and press SELECT button to start
device. Start stopwatch.
Attention: If device runs continuously at any time
during test (as determined by a quiet "putt-putt"
sound), seal of test equipment is inadequate and
test will provide an inaccurate result (alarm will not
activate).
Solution: Check canister for signs of cracks, check
O-ring for damage and ensure tethered cap is fully
inserted into canister tubing quick click connector.
4.
Verify within 5 minutes the screen displays "!!
WARNING !! BLOCKAGE/FULL", status/alarm
indicator illuminates solid yellow, and audible alarm
beeps 2 times every 20 seconds.
Note: Due to device normal internal leakage, device
may run at a very slow rate but still activate "!! WARNING
!! BLOCKAGE/FULL" alarm after approximately 5
minutes. If alarm is activated, device passed test
successfully.
5.
If device fails on first attempt check for possible
air leaks in system. Address leaks, and re-test. If
device fails a second time, remove from service and
return for repair.
6.
If alarm activates, press AUDIO PAUSE button. Verify
audible alarm is temporarily paused, the audio
pause indicator illuminates solid yellow, and screen
displays "!! AUDIO PAUSED !! BLOCKAGE/FULL".
7.
Press SELECT button to stop device, and pause
stopwatch.