Steps
Functionality check - Battery function check (performed after each patient use)
36
During the 2 seconds the POWER button is
pressed, status/alarm indicator illuminates solid
yellow and all indicator lights illuminate.
37
Display shows "Welcome, Starting", current
software version (e.g. V 1.06), and changes to show
"Standby" and therapy set-point.
38
Device starts operating, status/alarm indicator
illuminates solid green, and screen displays
"Continuous" and therapy set-point.
Alarm indicator check – Complete blockage / canister over capacity alarm and audio pause button check
(performed after each patient use)
39
Within 5 minutes the screen displays
"!! WARNING !! BLOCKAGE/FULL", status/alarm
indicator illuminates solid yellow, audible alarm
beeps 2 times every 20 seconds.
40
The audible alarm is temporarily paused, audio
pause indicator illuminates solid yellow, and screen
displays "!! AUDIO PAUSED !! BLOCKAGE/FULL".
Alarm indicator check – Low vacuum, leak and audio pause button check (performed after each patient use)
41
Within 30 ± 5 seconds the screen displays "
!! WARNING !! LOW VACUUM", status/alarm indi-
cator illuminates solid yellow, and audible alarm
beeps 2 times every 20 seconds.
42
The audible alarm is temporarily paused, audio
pause indicator illuminates solid yellow, and the
screen displays "!! AUDIO PAUSED !! LOW VACUUM".
43
Devices with software V 1.06 or later
Within 1 minute ± 5 seconds the audible alarm re-
sets and beeps 2 times every 20 seconds and the
screen displays "!! WARNING !! LEAK" alternating
with "!! WARNING !! LOW VACUUM".
Alarm indicator check – Over vacuum check (performed after each patient use)
44
Device stops, screen displays "!! THERAPY STOP !!
OVER VACUUM" alternating with "!! THERAPY STOP !!
Restart Device", status/alarm indicator illuminates
solid yellow, audible alarm beeps 2 times every
20 seconds and the keypad lock indicator
illuminates solid blue.
45
The audible alarm does not pause, and the screen
displays "!! Press !! POWER Key" while the AUDIO
PAUSE button is pressed.
46
Device does not restart, audible alarm does not
pause, and the screen displays "!! Press !!
POWER Key" while the SELECT button is pressed.
Yes No
Comments
17