Functionality check
Recommended personnel: Functionality checks should
be performed by a trained technician.
Equipment required: RENASYS™ power cord,
RENASYS power supply, RENASYS quick click adaptor,
RENASYS GO canister.
Operations check
1.
Connect power supply to AC mains power inlet on
device. Connect power cord to power supply and to
AC mains power.
2.
Verify all indicator lights and status/alarm indicator
flash briefly as AC mains power is connected.
3.
Verify battery indicator light flashes green, or
illuminates solid green if battery is fully charged.
4.
Press and hold POWER button for 2 seconds to turn
on device. Verify display screen shows "Welcome,
Starting", current software version (e.g., V 1.06), and
changes to show "Standby" and therapy set-point.
Record software version on record sheet.
Language check
To change language selection:
1.
Press and hold POWER button for 2 seconds to turn
device off.
2.
Simultaneously press and hold UP, SELECT and POWER
buttons for 2 seconds.
3.
Language can now be selected using UP and
DOWN buttons to scroll to desired language
followed by SELECT button.
4.
Screen will display "Welcome, Starting", current
software version (e.g. V 1.06), and change to show
"Standby" and therapy set-point.
Note: The language should be set back to
language setting upon device receipt once checks
are completed.
Modes of operation check
1.
Attach canister to device and engage orange
canister clips. Ensure canister viewing window is
facing forward.
2.
Connect canister quick click connector to quick
click adaptor and engage tethered cap into open
connecter of quick click adaptor.
Continuous mode:
1.
Adjust therapy set-point to 80mmHg using UP
and DOWN buttons, press SELECT button to start
device. Verify device starts operating, status/alarm
indicator illuminates solid green and screen displays
"Continuous" and therapy set-point.
2.
Press SELECT button to stop device operation. Verify
device stops operating, status/alarm indicator is no
longer illuminated, and screen displays "Standby"
and therapy set-point.
8