Steps
Electrical safety check (performed annually)
47
Power cord protective check resistance check
(perform test only for class I power supply REF
66800161):
Power cord acceptance criteria: electrical resistance
<100 mΩ
IN ACCORDANCE WITH IEC62353
48
Equipment leakage current check (normal polari-
ty): Acceptance criteria (class I power supply REF
66800161 paired with appropriate power cord):
Max: 500μA
Acceptance criteria (class II power supply REF
66801558 paired with appropriate power cord):
Max: 100μA
IN ACCORDANCE WITH IEC62353
49
Equipment leakage current check (reversed polar-
ity) Acceptance criteria (class I power supply REF
66800161 paired with appropriate power cord):
Max: 500μA
Acceptance criteria (class II power supply REF
66801558 paired with appropriate power cord):
Max: 100μA
IN ACCORDANCE WITH IEC62353
Device pressure check (performed annually)
50
Pressure measured on the calibrated pressure
gauge controls to 40mmHg ± 10mmHg.
51
Pressure measured on the calibrated pressure
gauge controls to 180mm ± 10mmHg.
Battery charge: Charge level check (preparation for return to clinical use)
52
Battery charge level is above 50%.
Clearing therapy time (preparation for return to clinical use)
53
Therapy time cleared.
Total time of operation (preparation for return to clinical use)
54
Total time of operation recorded.
Cleaning and packing
55
Device cleaned and re-packaged with all necessary
items for customer use.
Performed by:
Print Name:
Signature:
Location (Address):
Yes No
Comments
Class:
Class:
Class:
mmHg
mmHg
Date:
Day
18
μA
μA
Month
Year