6.
Snap the filter cover back into place over the filter.
7.
Plug the power cord into the back of the analyzer.
8.
Press the Power button. The analyzer then performs a diagnostic self test, and may take a
few minutes to bring the rotor chamber to operating temperature.
Note:
Refer to
test after cleaning the air filter. See
ning samples.
10.4 Updating the Analyzer Software
Abaxis provides software updates on CD-ROM to registered owners of the VetScan VS2 Analyzer.
Replacement software is mailed immediately whenever a new software version is released. See
ing Analyzer Identification" on page 5-13
Contact Abaxis Technical Support immediately at 1-800-822-2947 if the software version is not the
most recent release.
Note:
Be sure to update the software every time you receive an update CD from Abaxis.
CAUTION:
Updating software can return the VS2 to its default settings, and delete any stored
records and custom reference ranges. Before beginning the update, make sure you
archive your data. For instructions, see
ence Ranges," on page
You will also have a chance to cancel the procedure after beginning.
Contact Abaxis Technical Support for information on software updates and how they
can affect your VS2.
1.
In the Home screen, touch the Settings icon.
2.
Touch the More Settings icon.
3.
Touch the Software Update icon.
"Troubleshooting" on page 8-1
5-10.
Maintenance and Service
if the analyzer does not successfully pass the self
"Running Tests" on page 4-4
to determine the software version installed in the analyzer.
Section 5.3, "Printing and Archiving Refer-
for instructions on run-
"View-
10-3