iQC Code
iQC Message
Temperature Error
4042
iQC Bead Mix Error
4044
RQC Error
4045
Flash Lamp Level Error
4051
Engine Reset Error
4053
Sample Quality Error
4069
Bar Code Format Error
410D
410E
Bar Code Format Error
410F
4110
4111
Drawer Open Jam
4220
Drawer Closed Jam
4221
Note:
If the iQC code is not listed or if you cannot correct a particular problem, record the iQC
number or print the report, and contact Abaxis Technical Support, 24 hours a day, 7 days a
week, at 1-800-822-2947, or send a fax to 1-877-900-9333.
Solution
Temperature deviated more than one degree
during run. Accuracy of certain chemistries
may be affected. Rerun with new rotor.
Analyzer cannot determine exposure status
if iQC beads do not mix correctly. Possibly
triggered by lipemic sample. Spin and check
sample integrity. Contact Abaxis Technical
Support or verify integrity before rerun.
Reagent QC beads recovered below limit.
Rotor may have been stored incorrectly.
Rerun with new rotor.
Analyzer checked lamp (to make sure it
flashes when optical cuvette is sampled) but
detectors did not recognize lamp flashes. If
symptom persists, contact Abaxis Technical
Support.
Analyzer component restarted unexpectedly.
Contact Abaxis Technical Support.
Sample may contain physical interferents.
Confirm sample integrity and rerun rotor.
Analyzer could not read rotor's bar code. If
symptom persists, contact Abaxis Technical
Support.
Analyzer did not recognize rotor. Contact
Abaxis Technical Support.
Drawer cannot open correctly. Remove any
obstructions. If symptom persists, contact
Abaxis Technical Support.
Drawer cannot close correctly. Remove any
obstructions. If symptom persists, contact
Abaxis Technical Support.
iQC Codes
A-3