3.
Complete the warranty online at www.abaxis.com/warranty within 10 days of installation to
start the warranty period. Customers are placed on the customer mailing list to receive infor-
mation pertaining to the VS2 and ancillary products, such as software upgrades and new prod-
uct announcements.
3.2

Setting Up the Analyzer

Refer to the required environmental specifications in
page
3-5.
1.
Make sure that the VetScan VS2 is:
On a sturdy, level surface, free of vibration and sudden jolts.
In an ambient operating temperature of 15–32 ºC (59–90 ºF).
In an environment free of animal hair, dust, and other contaminants.
Not placed near a sunny window or another heat source.
At least six inches from the wall for access to the power connection and USB ports.
Note:
If the analyzer is subjected to an electrostatic discharge event, you may need to restart
the unit.
2.
Plug the power jack into the analyzer, and
plug the detachable power supply cord into
the power adapter and into a grounded elec-
trical outlet.
CAUTION:
To prevent power surges or drain, do
not plug the analyzer into the same
circuit as a centrifuge or any other
high-current device. If this is not pos-
sible, use an ancillary surge protector
or battery backup for the analyzer.
WARNING: U
SE ONLY THE
A
NY OTHER POWER SUPPLY WILL
DAMAGE THE INSTRUMENT AND VOID
THE WARRANTY
3-2
VS2
POWER SUPPLY
.
Set Up and Analyzer Description
"Analyzer and Environmental Specifications" on
.