No
Go to the next step.
4. The patient brought only the patient ID card to the MRI appointment, and the
neurostimulator model number on the card is listed on page 5.
Confirm that the neurostimulator stimulation parameters are set as follows:
Stimulation:
Other parameters:
In addition for the Itrel 3 Model 7425 only:
Magnetic (reed) switch:
Yes
Yes, confirmed. Go to "Head-only eligible – MRI equipment and scan
requirements" on page 20 and continue from there.
No
No, could not confirm. Contact Medtronic Technical Services.
If all of the instructions stated in this head-only eligible section are followed, MRI
scans of the head only using an RF transmit/receive head coil may be safely
performed.
Off
If you are not certain if stimulation is off, ask the patient if
it is off.
No change
Disabled
MRI guidelines for Medtronic neurostimulation systems for chronic pain English 19