Analog Output Connections

The analog output option board (Figure 2-35) fits into the PLC interface slot on the
Main board. It provides either 0-10 VDC or 4-20mA (only one) analog signal
proportional to the weight applied to the scale.
Figure 2-35: Analog Output Option Board
Connections to this option should be made as shown in Figure 2-36.
Figure 2-36: Wiring the Analog Output
IND560 PDX Installation Manual
NOTES:
1. USE TWO-CONDUCTOR SHIELDED CABLE.
2. MINIMUM RESISTANCE OF DEVICE LOAD: 500
OHMS.
2
3. WIRE SIZE: 18 AWG (.823 mm
) MAXIMUM
2
24 AWG (0.205 mm
) MINIMUM.
2-23