Connecting
3.4 Programming the Transmitter
Table 3- 3
Pipe Configuration Option List Definitions
Options
Fully Developed
Fully developed flow, as would be expected for very long straight pipe runs or installation downstream
of a flow condition.
1 Elbow
Single 90 degree Elbow upstream of sensor installation.
Dble Elbow+
Double out-of-plane Elbows upstream of sensor installation.
Dble Elbow-
Double in-plane Elbows upstream of sensor installation.
Valve
Not available at this time.
Expander
Pipe expansion upstream of sensor installation.
Reducer
Pipe reduction upstream of sensor installation.
Norm Entry
Not available at this time.
Header Inlet
Header or pipe manifold upstream of sensor installation.
Intrusions
Not available at this time.
Selecting Sensor
The following is a typical sensor selection procedure for D1H High Precision Sensors.
1. From the [Channel Setup] menu press the <Down Arrow> to select [install Xdcr].
2. Press the <Right Arrow> to [Transducer Model]. Press <Right Arrow> and scroll to
3. To select [Transducer Size], press <Right Arrow> and select [D1H]. Press <ENT>.
4. IMPORTANT: Record Number Index and Spacing Method.
28
[1011HP-T1]. Press <ENT>.
Select based on type.
Select based on size
After sensor is mounted select "Install.
Definitions
Operating Instructions, 7/2010, A5E03086486-02
FUS1010 IP65 NEMA 4X