Figure 3-9
Clamp-on style sensors are mounted on the surface of the monitored pipe using series
mounting assemblies. Apply a generous quantity of the thermal couplant provided to the tip
of the sensor and attach it securely to the cleaned pipe surface with the proper mounting
assembly. Temperature measurement anomalies resulting from variations in the ambient
conditions can be minimized by insulating the pipe and sensor after installation.
FUS1010 IP65 NEMA 4X
Operating Instructions, 7/2010, A5E03086486-02
992EC Series Cable
Clamp-on Temperature Sensor (Matched
pair required)
Clamp-on Temperature Sensor (only FUE1010)
Thermal Couplant
Pipe
Mounting Assembly
Connecting
3.5 Sensor Installation
35