3.4
Suction handpieces
3
59 06 511 D 3386
D 3386.103.01.06.02
5
2
1
4
3
Sterilization
The suction cannula (1),
the intermediate piece (2),
the suction handpiece (3),
the saliva ejector handpiece (4)
and the surgical suction cannula (5)
can be sterilized.
Sterilize only in an autoclave at 135°C, 2.1 bar.
Thermodisinfection
All parts can be thermodisinfected at 93°C (197°F).
Disinfection
All parts can be sprayed or wiped with a disinfecting
solution.
ATTENTION
Use only cleaning agents which have been approved by
Sirona !
Grease the O-rings at the joints afterwards!
Weekly
Separate the suction handpiece at the joint, clean the
handpiece, and grease the joints.
ATTENTION
Use only lubricants which have been approved by
Sirona!
3 Vacuum system
3.4 Suction handpieces
11