3 Vacuum system

3.3 Disinfection when using wet aspiration

3.3
Disinfection when using wet aspiration
10
simultaneously
With this system, the automatic separating device and,
where appropriate, the amalgam separator are not
integrated into the water unit. Both separation processes
are performed centrally.
Daily before the mid-day break and
after the practice has closed
• Flush through the vacuum system as described under
3.2 Disinfection for dry aspiration with a suitable
cleaning and disinfection agent.
• After the liquid has been aspirated through the two
suction hoses, remove the cleaning adapter.
• After this, pour 1 liter of disinfectant solution into the
cuspidor while simultaneously pressing the button
(arrow) for approx. 10 seconds.
Pressing the button enables the cleaning and disinfec-
tant liquid from the cuspidor to reach the vacuum line
via the cuspidor valve.
• Finally, refit the gold trap.
ATTENTION
Use only cleaning and disinfecting agents which have
been approved by Sirona!
Do not use any detergents or other disinfectants!