Sirona Dental Systems GmbH
5
Instrument hoses
5.1
Removal/replacement of instrument hoses
5.2
Turbine hose
59 06 511 D 3386
D 3386.103.01.06.02
5.1 Removal/replacement of instrument hoses
First switch the unit OFF!
Unscrew the cap nut (1) and then simply pull off the
instrument supply hoses.
i
NOTE
Only SIROTOM, SIROSONIC L, POLYLUX or
SIROCAM 3 or SIROCAM C may be connected to the
right coupling piece.
Only SIROSONIC L or a 4th drill drive (only with
expansion set) may be connected to the second coupling
piece from the right.
At the unit end of the turbine hose, return air leaks out
with a small amount of turbine oil. This oil is absorbed by
the cotton wool roll in the transparent collecting cup.
We recommend replacing the cotton wool roll every
week.
To do so, push the cup down and remove the cotton
wool roll.
Cut off and insert a new cotton wool roll.
Push the cup back into place.
5 Instrument hoses
15