Sirona Dental Systems GmbH
San.
Sanier – Phase
Stop
(Sanitation phase)
1
2
3
4
San.
Sanier – Phase
Stop
(Sanitation phase)
1
2
3
4
Sanier – Phase
(Sanitation phase)
1
2
3
4
59 06 511 D 3386
D 3386.103.01.06.02
24:00
1
2
5
3
4
5
6
1
2
6
3
4
5
6
1
2
1
3
4
5
6
Sanitation phase 5
If all number symbols of the instruments removed for
sanitation are dark, the reaction time of 24:00 hours
appears on the top of the display.
• The unit must now be switched OFF for at least 24
hours, but not longer than three days.
Leave all hoses plugged unchanged in the adapters of
the sanitation hood.
• Switch the unit back ON after a minimum of 24 hours.
i
NOTE
If you have blocked the water supply for the treatment
center after sanitation phase 5, the water supply must be
opened again prior to renewed start-up of the treatment
center (i.e. after 24 hours). Switch the unit back ON after
this.
Leave all hoses plugged into the adapters of the
sanitation hood. The sanitation program then
automatically runs to its completion.
If you switch on the treatment center before the 24 hours
have elapsed (the missing time is then indicated on the
display), you must activate the "San. Stop" key after
switching on the unit. The sanitation program then runs
to its completion.
Sanitation phase 6
The hoses and the tumbler filler are rinsed with water.
This takes several minutes.
The receptacle symbols are again marked with a bold
frame after rinsing.
• As soon as sanitation phase 1 appears again on the
display, the hoses can be placed in the corresponding
receptacles. The "Back" arrow then appears at the top
left on the display.
• Now activate the "Back" arrow to leave the sanitation
dialog.
The unit is now ready for operation again.
6 Disinfection system
6.2 Sanitation
21