4 Amalgam separation
3.6 Thermodisinfecting the suction hoses
4
Amalgam separation
Start
San.
Amalgam
6
5
7
loosen
tighten
Replacement
amalgam rotor
Order No.:
14 34 138
2
7
14
0m
42s
C
12:30
2
1
4
O-ring
3
Grease
3
Disposal
When the message Amalgam lights up on the display of
the dentist element, proceed as follows:
• Disinfect as described in chapter 3 "Vacuum sys-
tem".
• Pull the lower cover (1) off from the water unit.
DO NOT turn the unit off!
• Turn the lower part (2) of the amalgam separator
with rotor (3) in the direction of the arrow until it can be
removed easily.
The message Amalgam flashes on the displays.
• Put any amalgam residues from the suction hoses and
the cuspidor into the dismantled rotor for disposal (see
pages 7 and 12).
• Hold the bottom part upright and click on the trans-
3
port cap (4). This releases a binding agent. Therefore,
do not remove the transport cap again!
• Remove the rotor and cap from the lower part and
place it in the special packing for disposal.
(Send it to one of the companies listed on the insert
enclosed with the new amalgam rotor
• Insert the new rotor (3) in the lower part (2).
Do not install a used rotor!
• Reinstall the lower part and turn it in the opposite di-
rection of the arrow up to the stop. The bar (7) should
point to the chair at the start and to the siphon (6) after
it is screwed tight.
If after insertion of the rotor Amalgam still flashes on
the displays, then the lower part is not correctly locked
in place!
• Refit the cover (1).
NOTE: Independently of the AMALGAM message
lighting up, the rotor (3) in the amalgam separator must
be replaced at least once per year, otherwise
unpleasant odors can develop.
Siphon
Clean the siphon (6) in case of poor drainage of the
cuspidor, otherwise once a month.
To do so, first open the cover (1) and unscrew the cap
(5).
Have a collector ready to catch the contents!
Reinstall the cap and cover.
i
NOTE
Always wear gloves when performing these tasks!