6 Disinfection system
6.2 Sanitation
Start
San.
Sanier – Phase
(Sanitation phase)
1
2
3
4
3
4
1
2
18
0m
42s
C
12:30
1
2
1
3
4
5
6
3
4
1
Activating the sanitation
• In the Chair program dialog select the assistant di-
alog.
• In the assistant dialog, activate the San. key for ap-
prox. 3 seconds.
Activation is only possible if there is sufficient
DENTOSEPT P in the refill container and if all
instruments are fitted firmly in their instrument
receptacles.
If necessary refill DENTOSEPT P and check the correct
fit of the instruments.
The sanitation dialog appears.
Sanitation phase 1
During the sanitation procedure, the displays indicate
sanitation phases 1 to 6.
The left key row identifies the receptacles on the
assistant element.
The right key row identifies the receptacles on the
dentist element.
Preparation of the sanitation hood
• Place the sanitation hood with lid on the cuspidor. The
sanitation hood vessel must always be clean!
• Insert the cleaning adapters for the instrument cou-
plings (1), SPRAYVIT valve bodies (2) and suction ho-
ses (3) up to the stop in the sanitation hood, as shown.
The adapters for the instrument couplings are color-
coded to match the hose couplings.
SL motor: green
Electric motors with interface acc. to ISO 3964 such as
2
EL 1 motor: blue
Turbine: white
SIROSONIC L: red
• For the saliva ejector hose, insert the intermediate pi-
ece (4) in the cleaning adapter.
i
NOTE
To prevent the hoses from twisting and stretching, the
arrangement of the cleaning adapters from left to right
should correspond to the arrangement of the instruments
on the dentist element.
Sirona Dental Systems GmbH
59 06 511 D 3386
D 3386.103.01.06.02