7. Reconnect the battery power connector to the main board (Figure 2-8).
2.3.2.3.
POWERCELL Model, PDX or GDD
64084463
|
03
|
12/2017
1. Open the enclosure as detailed in the Opening the Enclosure section. For steps 3 to 5, refer
to Figure 2-10.
2. First, the orientation of the front panel must be reversed. This involves detaching the panel
and rotating the housing, and exchanging the openings used by the power and load cell
cables:
Figure 2-10: Preparing to Detach Front Panel
3. Loosen and remove the two nuts securing the two hinge/grounding straps that connect the
front panel to the rear housing.
4. Cut the nylon wire tie securing the AC wires to the plastic adhesive pad.
5. Carefully rotate the housing 180 degrees and use the nuts removed in step 4 to reattach
the two grounding straps to the studs near the grip bushings (Figure 2-11). Tighten the two
nuts.
METTLER TOLEDO IND246 and IND246 POWERCELL Installation Manual
Figure 2-9: Housing Reversed
Housing, rotated
Ground strap connection studs
Ground wire connected
to ground stud
Incoming power cord
Attachment studs
Power connector
2-7