Active Input
Selecting the inputs as active (Figure 2-70) enables connection of switches or other simple devices
to trigger an input. No voltage is supplied by the external simple device.
An example of how to wire to the active inputs is shown in Figure 2-63.
2.4.6.3.3.
Passive Input
Selecting the inputs as passive (Figure 2-70) enables other devices such as PLCs to provide the
trigger voltage (typically 12 VDC or 24 VDC, maximum 30 VDC) to turn the IND246 inputs "on".
An example of wiring to the passive inputs with the +V to the common is shown in Figure 2-64.
64084463
|
03
|
12/2017
NOTES:
1. VOLTAGE IS LOGIC LEVEL 5 VDC. LOW RESISTANCE
CONTACTS RECOMMENDED. MAXIMUM CABLE LENGTH:
20 FEET / 6 METERS.
2.
DO NOT BUNDLE INPUT WIRING WITH POWER WIRING OR
ANY OTHER HIGH ENERGY CABLES.
3.
SWITCHES MAY BE REPLACED WITH RELAY DRY
CONTACTS.
4.
WIRE SIZE: 18 AWG (0.832 mm
Figure 2-66: Active Input Connections
Figure 2-67: Passive Input Connections
METTLER TOLEDO IND246 and IND246 POWERCELL Installation Manual
2
) MAXIMUM
2
24 AWG (0.205 mm
) MINIMUM
NOTES:
1. VOLTAGE: 5-30 VDC, 10mA MAXIMUM CURRENT.
VOLTAGE POLARITY MAY NOT BE REVERSED.
2. DO NOT BUNDLE INPUT WIRING WITH POWER WIRING OR
ANY OTHER HIGH ENERGY CABLES.
2
3. WIRE SIZE: 18 AWG (0.832 mm
) MAXIMUM
2
24 AWG (0.205 mm
) MINIMUM