2.4.4.3.
External Grounding of the Cable
1. Disassemble the external grounding feature provided with the IND246 POWERCELL terminal – it
2. Slide the outer drain wire through the clamp, next to the flat braided cable.
3. Slide the end of the mounting bracket into the clamp.
4. Fasten the mounting bracket to the enclosure using the screw at the external grounding point
5. Adjust the lengths of the braided cable and outer drain wire as necessary, and tighten the
6. Wrap the free end of the braided grounding cable about two turns around the POWERCELL
2-26
METTLER TOLEDO IND246 and IND246 POWERCELL Installation Manual
includes a screw clamp and a mounting bracket. Slide the screw clamp about 8"/20 cm onto
the flat braided grounding cable.
Figure 2-48: Grounding Feature and Flat Braided Cable
Figure 2-49: Outer Drain Wire Inserted Through Grounding Clamp
Figure 2-50: Grounding Clamp Mounting Bracket Inserted
indicated in Figure 2-27.
clamp on the grounding feature.
cable. Adjust the length of the braided cable as necessary.
About 8"/20 cm
of cable
64084463
|
03
|
12/2017