Sterilization Validation Specifications
EtO sterilization validation for this product was performed at the parameters listed below. Use
these or equivalent parameters in your EtO sterilization process.
Preconditioning
Vacuum:
Humidity:
Temperature:
Time:
Exposure Cycle
EtO Concentration:
Humidity:
Temperature:
Concentration:
Exposure Time:
Intentional leak
Typical Exhaust Flow / Pressure Performance of the Whisper Swivel II
45
40
35
30
25
Flow
20
(SLPM)
15
10
5
0
0
Disposal
Dispose of in accordance with local regulations.
Murrysville, PA 15668 USA
21" Hg (± 1 Hg)
70% (± 10%)
54° C (± 2°)
60 minutes
12% (± 1%)
70% (± 10%)
54° C (± 2°); 129.2° F(±3.6°)
600 mg/L (± 50 mg/L)
120 minutes
5
10
15
Pressure (cm H O)
2
20
25
© 2012 Koninklijke Philips Electronics n.V. All rights reserved.
1067701 R06
CV 03/05/2012