CAUTION: DO NOT use the device until it is completely dry
(Fig. 5f). Inhalation of disinfectant fumes can cause severe
coughing and irritation of the airways.
If the Provox LaryButton looks dirty or has air dried in an
area with a risk of contamination the device should be both
cleaned and disinfected before use. A risk of contamination
could exist if the device has been dropped on the floor or if
it has been in contact with a pet, someone with a respiratory
infection, or any other gross contamination.
During hospitalization it is important to both clean and disinfect
the Provox LaryButton after use but also before use since there
is an increased risk of infection and contamination. In a hospital
it is better to use sterile water for cleaning and rinsing, rather
than drinking water.
CAUTION: DO NOT clean or disinfect by any other method
than described above as this might cause product damage and
patient injury.
2.4 Storage instructions
When not in use, clean and disinfect the device as described
above and then store the Provox LaryButton in a clean and dry
container at room temperature. Protect from direct sunlight.
2.5 Device lifetime
The Provox LaryButton may be used for a maximum period
of 6 months. If the device shows any signs of damage it shall
be replaced earlier.
2.6 Disposal
Always follow medical practice and national requirements
regarding biohazard when disposing of a used medical device.