Device removal
1. Hold the Provox LaryButton in place and gently remove
the Provox HME System component from the HME Holder
(Fig. 4a).
2. Grasp the Provox LaryButton and gently turn it (about half
a turn) to allow the Retention Collar to collapse thereby
reducing its width (Fig. 4b).
3. Remove the device by gently tilting it out of the stoma (Fig. 4c).
Alternatively, the device can be removed by pushing with a
finger on the shaft from one direction and tilt the device out.
NOTE: The insertion or removal procedures may occasionally
cause some slight bleeding, irritation or coughing. Inform your
clinician if the bleeding persists.

2.3 Cleaning and disinfection

CAUTION: Always remove the device from the stoma before
cleaning.
DO NOT at any time attempt to clean the device while inside
the stoma since this could cause injury.
Clean the Provox LaryButton after each use, or at least daily
with the Provox TubeBrush (Fig. 5a-5g).
Do not use any water other than drinking water to clean
and rinse the device.
CAUTION: Be careful not to burn your finger in the warm
cleaning water.
Disinfect the Provox LaryButton at least once a day (Fig. 5e).
with one of the following methods:
• Ethanol 70% for 10 minutes
• Isopropylalcohol 70% for 10 minutes
• Hydrogenperoxide 3% for 60 minutes