Descargar Imprimir esta página

Bipolar Mode Autostart; Bipolar Mode Autostop; Activation Using The (Monopolar) 3-Pin Handpiece - Storz AUTOCON II 400 Manual De Servicio

Publicidad

4.3.10 Bipolar mode autostart

– Prepare the testing equipment test box autostart / autostop.
– Connect the bipolar cable to connections no. 1 and no. 2 on the autostart / autostop test box.
– Select the bipolar socket.
– In the "Select Activate Type" menu, activate autostart 1 mode as the footswitch function.
– Proceed as follows to set the operating mode:
Bipolar Soft Coag
Power: 50 W
Effect: 1
– Press button S2 on the autostart / autostop test box and hold it down. At the end of autostart time 1
the device will be activated (its default value of 1 sec. can be changed if necessary in the "Device
Setup menu: Application Settings").
– Proceed as follows to set the operating mode:
Bipolar Soft Coag
Power: 50 W
Effect: 8
– Press button S2 on the autostart / autostop test box and hold it down. At the end of autostart time 1
the device will be activated (default value: 1 sec.).
– Change the activation mode from autostart to footswitch.

4.3.11 Bipolar mode autostop

– Connect the bipolar cable to connections no. 3 and no. 4 on the autostart / autostop test box.
– Connect the dual-pedal footswitch.
– Make the dual-pedal footswitch the active footswitch.
– Proceed as follows to set the operating mode:
Bi-Soft Autostop
Power: 50 W
Effect: 4
– Press button S3 on the autostart / autostop test box.
– Activate the "Coagulation" footswitch.
– Press button S3, hold it down for approx. 4 sec. and then release it.
– The device must continue to be activated for a further 4 to 10 sec. The device must end activation
automatically. An acoustic message "Autostop reached" will be sounded.
– Keep the footswitch depressed for 10 sec. more. After 5 more seconds of activation, the alarm
message "A-10: Release activation (Termination by Autostop)" should appear. The message is
displayed on the screen against a turquoise background.
– Deactivate the footswitch.
– Leave the device switched on.
4.3.12 Activation using the (monopolar) 3-pin
handpiece
Note:
This test item is not applicable to device versions –113 or –114 (but does also apply to Uxxy
versions).
Note:
The user's neutral electrodes and NE cable should be used as a rule. Alternatively you can use
your own measuring equipment. Use handpieces from manufacturers of disposable products
(e.g. mtp 010103-40) – this enables diode errors in the sockets to be detected.
– Connect the monopolar handpiece to the third HF socket (monopolar 3-pin).
– In the "Device Setup: Application Settings" menu, set NE type to "any".
– From the procedure list, select a procedure with a monopolar application or, if necessary, create a
suitable procedure.
– As described above, plug the neutral electrode cable into the NE socket. The NE indicator on the
device should light up green.
– Select the "Monopolar Socket" button. A new window will open.
4-8
AUTOCON
®
II 400
MODEL 205352 20-1
12.06 / V1.0

Hide quick links:

Publicidad

Capítulos

Solución de problemas

loading