Error code
Description of error
A-95
Instrument recognized at MF socket
A-98
CRC error at RAM self test
A-9A
Incompatible MF instrument
A-9F
MF instrument disconnected
Missing status message from HF
A-A1
module
A-A3
Footswitch is not assigned to a socket
A-A4
Two identical footswitches connected
A-A9
Select procedure prior to activation
A-AA
No mode assigned
A-AB
No/Wrong MF instrument detected
Autostart cannot be assigned at tissue
A-AC
contact
A-AF
NE monitor ceases to respond
A-B0
Bad NE symmetry
A-B1
NE current density too high
A-B7
Power is limited to 50 W with autostart
A-BB
Next safety check date overdue
No activation type for MF socket
A-C0
selected
A-C1
Activation not allowed
A-C2
Defective front panel fan
A-C3
Select new mode for MF instrument
A-FB
Socket configuration has been changed
A-FE
Power fail signal detected
A-FF
Invalid password
C-01
Missing ESI message from front panel
C-04
Internal status error
C-05
Wrong ECB message length
C-06
No parameter received
C-0D
Wrong ECB message length
C-21
Erroneous power dosage
C-26
Erroneous power measurement
C-30
HF voltage too high (redundant)
C-31
HF voltage too high (OFF)
C-32
HF voltage too low (OFF)
C-33
HF voltage too high (ON)
3-8
Solution
New MF instrument at socket
MF instrument disconnected from
socket
Assign footswitch to a socket
Either connect one footswitch or two of
different types
Select any procedure
Assign a mode to the procedure
Attempt activation at a socket at which
either no instrument or an invalid
instrument has been detected
Break tissue contact, disconnect
sections of the instrument as required,
assign autostart
Check neutral electrode and reposition
if necessary
The max. output with autostart should
only be 50 W
Agree date for safety check
Specify activation type
Check power cord, power outlet and
power supply
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
AUTOCON
®
II 400
MODEL 205352 20-1
Priority
Low
Medium
Low
Low
High
Low
Low
Low
Low
Medium
Medium
High
Low
Medium
Low
Medium
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
12.06 / V1.0