Descargar Imprimir esta página

Storz AUTOCON II 400 Manual De Servicio página 112

Publicidad

92
Technische Beschreibung
Technische Daten
FORCED Coag
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
HF power limitation
Max. power output at
rated load resistance
SPRAY Coag
HF voltage form
Rated frequency
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
HF power limitation
Max. power output at
rated load resistance
BIPOLAR SOFT Coag / BIPOLAR SOFT with AUTOSTOP
HF voltage form
Rated frequency 350 kHz
Crest factor
Rated load resistance
Max. HF peak voltage
Number of effects
Constancy of effects
HF power limitation
Max. power output at
rated load resistance
Technical description
Technical data
pulse-modulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
6.0 (at RL = 500 ohms)
500 ohms
1800 Vp
4
automatic control of HF peak voltage
5 W to 120 W in 1 W steps
120 W ± 20%
pulse-modulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
7.4 (at RL = 500 ohms)
500 ohms
4300 Vp
2
limitation of HF peak voltage
5 W to 120 W in 1 W steps
120 W ± 20%
unmodulated sinusoidal AC voltage
350 kHz (at RL = 500 ohms) ± 10%
1.4 (at RL = 500 ohms)
75 ohms
190 Vp
8
automatic control of HF peak voltage
1 W to 120 W in 1 W steps
120 W ± 20%