Descargar Imprimir esta página

Storz AUTOCON II 400 Manual De Servicio página 201

Publicidad

Test report – safety checks
Article number
Device type
Serial number
Software versions
Generator control
ESI
User interface
SCB
Visual inspections
Labeling
Potential equalization
Fuse type (T 8.0 AH / T 4.0 AH)
Adhesive label
Fuse labels
Serial number
Manufacturer's identification plate
CE mark
TÜV / NRTL
Components
Feedthrough cover
Power switch
green when on
Socket modules
undamaged
Frame
undamaged
Housing
undamaged
Electrical safety
Protective ground
resistance (with
power cord)
Earth leakage
N.C.:
current
S.F.C.:
Housing leakage
N.C.:
current
S.F.C.:
Patient leakage
N.C.:
current
S.F.C.:
Patient leakage
current (with line
S.F.C.:
voltage on applied
part)
Protective conductor testing
Protective conductor contact, power
socket against housing
Housing base
Cover
Front panel
Front frame
PE contact feedthrough (3-pole)
Bus termination testing SCB
Socket 1
one plug connected
both plugs
Socket 1
connected
Socket 2
one plug connected
both plugs
Socket 2
connected
2) Tested only when the device is linked up to KARL STORZ-SCB
3) Tested only when a triple-pedal footswitch is installed.
4) When a handpiece 205300 43 and high-frequency cable 205300 45 are used.
12.06 / V1.0
0.20
0.50 mA
1.00 mA
0.10 mA
0.50 mA
0.01 mA
0.05 mA
0.05 mA
2)
Device startup
Procedure list
Alive signal (clock)
Footswitch
Message detected
recognition
Message separated
Activation signal
Bipolar Cut
Bipolar Coag
3)
Procedure switchover A / B
Only triple-pedal footswitch
4)
Handpiece detection
Activation signal
Cut
Coag
Device startup – Error detection
Touch screen
Error message
pressed
Footswitch
Error message
pressed
Handpiece
Error message
pressed
NE monitoring
Limit resistance
single-faced
Limit resistance
double-faced
Symmetry
double-faced
Autostart bipolar
Autostop bipolar
Instrument detection MF socket
Power measurement cut monopolar
Power Cut
300 W ±20%
Top Cut
300 W ±20%
C-Cut
200 W ±20%
Power measurement cut bipolar
Bipolar Cut
100 W ±20%
Saline C-Cut (MF)
370 W +9% / -20%
Power measurement, monopolar
coagulation
Standard Coag
200 W ±20%
Forced Coag
120 W ±20%
Spray Coag
120 W ±20%
Power measurement, bipolar
coagulation
Bipolar Soft Coag
120 W ±20%
Saline Coag (MF)
Date for next safety check
®
.
AUTOCON
®
II 400
MODEL 205352 20-1

Hide quick links:

Publicidad

Capítulos

Solución de problemas

loading