9.
Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Product not running
Product becomes too hot
Insufficient power
Loud running noise
Lid cannot be
attached/removed
Battery cannot be
inserted/removed
Sterile funnel or battery
removal device not
attachable
Speed control knob not
working
Cause
No battery
Battery not charged
Battery defective
Product in locked OFF position Safety catch is in OFF position
Product defective
Excessive use
Reprocessing/maintenance
carried out incorrectly
Fall damage, product defective Product heating
Blunt tool
Product defective
Blunt tool
Defective product drive/ball
bearings
Lid not compatible
Lid deformed/defective
Connection to product
deformed/defective
Battery not compatible
Battery deformed/defective
Battery shaft to product
deformed/defective
Sterile funnel or battery
removal device not compatible
Sterile funnel or battery
removal device
deformed/defective
Battery shaft to product
deformed/defective
Product in locked OFF position Safety catch is in OFF position
Speed control knob sticks/is
defective
Detection
No battery in the battery shaft
No signal tone when inserting the bat-
tery
No signal tone when inserting the bat-
tery
Product not running
Product heating
Product heating
Heating of tool and product
Insufficient product performance
Large amount of heat generated after a
short time
Tool blades worn out
Loud, perceptible noise during operation Follow operating instructions (preparation,
Lid does not click into place
Lid is difficult or cannot be
mounted/removed
Lid is difficult or cannot be
mounted/removed
Battery cannot be completely inserted
into the battery shaft
Battery is difficult or cannot be
mounted/removed
Battery is difficult or cannot be
mounted/removed
Sterile funnel or battery removal device
not attachable to the battery shaft
Sterile funnel or battery removal device
difficult or not attachable
Sterile funnel or battery removal device
difficult or not attachable
Speed control knob not working
Remedy
Insert battery.
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Follow operating instructions (operating mode).
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Change tool.
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Follow operating instructions (operating mode).
Have the manufacturer repair the product
Change tool.
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product
Use the compatible lid for GA334.
Have the manufacturer repair the lid.
Have the manufacturer repair the product.
Use the compatible battery for GA334.
Have the manufacturer repair the battery.
Have the manufacturer repair the product
Use compatible sterile funnel or battery removal
device for GA334.
Change sterile funnel or battery device.
Have the manufacturer repair the product.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
13