Net service
Connection
Network enabled: Activate by clicking the circle. Click again to de-activate.
WiFi name: Select your wifi network and enter your wifi password.
Net backup
Auto backup: Activate by clicking the circle. Click again to de-activate.
Manage backup: Export and import of data.
New backup password: Press OK. Enter new password - press OK.

9) TestPeg

The Osstell TestPeg may be used for testing and learning how to use the system. Place the Osstell TestPeg on the
table or hold it in your hand. Turn on the instrument, select a tooth position and hold the probe tip approximately
3-5 mm away from the top of the Osstell TestPeg. The Osstell IDx should start to measure and present an ISQ value
of 55 +/- 2 ISQ.

10) SmartPeg™

The SmartPeg is available with different connection geometries to fit all major implant products on the market.
You can find all available SmartPegs on osstell.com/smartpegguide.
The SmartPegs are disposable and should only be used for one or multiple measurements at one treatment
session, for a single patient only (to avoid cross-contamination). Repeated re-use may result in false readings due
to wear and tear of the soft aluminium SmartPeg threads.

11) How to Measure

A first measurement should be taken at implant placement to get a baseline for future measurements throughout
the healing process. Before the final restoration, a measurement is taken which makes it possible to observe the
stability development of the implant. We recommend you studying the more detailed information (videos and
quickguides) available on osstell.com/get-started, to utilize the full functionality of your IDx.
1.
Connect the probe to the instrument.
2.
Start the instrument
3.
Press and hold a tooth to prepare site.
4.
Enter data about the patient. We recommend using the same patient number as you
use in your own patient management system .
5.
Place the SmartPeg into the SmartPeg Mount
SmartPeg Mount will hold the SmartPeg. Attach the SmartPeg to the implant or abut-
ment by screwing the SmartPeg Mount using finger force of approximately 4-6 Ncm
Do not over-tighten, to avoid damaging the SmartPeg threads.
6.
Hold the probe close (3-5 mm) to the top of the SmartPeg without touching it
Buccal-Lingual
4
indicates that a measurement has been performed and the value, time and date are shown on the
display. Measured values can be deleted by clicking the trash can. Note that values cannot be deleted
after saving.
3
approx. 45º
3-5 mm
25070-05 ML
.
and Mesial-Distal direction
4
1
. The SmartPeg i magnetic, and the
5
. Do not measure from above
5
English
1
2
2
.
4-6 Ncm
3
. Measure in both the
6
. An audible sound
6