16) Interpret the Result

Implant Stability
An implant can have different stability in different directions. The stability of the implant is dependent of the
surrounding bone configuration. There is often a direction where the stability is lowest, and a direction where the
stability is highest and these two directions are most often perpendicular to each other.
To find the lowest stability (lowest ISQ value) it is recommended to measure from two different directions. The
lowest stability is in most cases found in the buccal-lingual direction. The highest stability is in most cases found
in the mesial-distal direction.
The ISQ Value
Assuming there is access to the implant, ISQ measurements should be performed at implant placement and be-
fore the implant is loaded or the abutment is connected. After each measurement, the ISQ values are used as the
baseline for the next measurement performed. A change in the ISQ value reflects a change in implant stability. In
general, an increase in ISQ values from one measurement time to the next indicates a progression towards higher
implant stability while a decrease in ISQ values indicates a loss in stability and, possibly, implant failure. A stable
ISQ value would indicate no change in stability.
The ISQ scale clinical guidelines are available by clicking on this symbol on the IDx screen.

17) Data Connection to Osstell Connect

Osstell Connect (osstellconnect.com) is an online service for data transfer, storage, display and overview of your
data. You can have your Osstell IDx connected to Osstell Connect by WIFI or ethernet cable.
Before you can start using data connection to Osstell Connect you have to register your Osstell IDx. The serial
number can be found on the back of the instrument. For registration assistance and features of data connection
to Osstell Connect, please visit: osstell.com/get-started.

18) Lock the unit

Press the on/off key
to lock the instrument. To unlock the unit: press the on/off key and enter the PIN code.

19) Cleaning and Maintenance

Before each use, parts in contact with the patient (Probe, Probe cable and SmartPeg Mount) should be cleaned
and sterilized by autoclaving. The instrument does not require regular maintenance. In the event of an instrument
malfunction contact the local sales representative or distributor for further instructions.
Instrument:
Can be cleaned using a damp cloth with water or isopropyl alcohol solution.
Do not use other cleaning fluids as they can damage the top cover of the instrument.
SmartPegs:
Single use only.
TestPeg:
Is not used intra-orally, does not require sterilization.
Probe and Smart-
Must be autoclaved according to the recommended sterilization method, validated to
Peg Mount:
sterility assurance levels (SAL), according to ISO 17665-1 and ISO 17664.
The Probe and SmartPeg Mount should be placed in suitable packaging for the steriliza-
tion process. Loosely coil the cable to allow for packaging without damage to the cable.
25070-05 ML
English
33 33