Battery Bay
Place the Charger on a flat, level surface away from sources of heat and moisture.
1.
Connect the Power Supply to the mains AC supply using the Cord provided.
2.
NOTE For the Two-bay Battery Charger (REF 00992020002) only, plug the DC connector from the Power Supply into the back of the
Charger.
Place the Battery in the Charger bay located on the
3.
Align the arrows of the Battery and the Charger to ensure that the Battery makes contact with the Charger
4.
5.
The LED status window on the Battery Charger will indicate the status of the Battery (refer to the LED indication section for details) and
the Charger will automatically begin charging.
Remove the Battery when the Charger LED status window indicator is green solid.
6.
Battery Charger
Arrow Alignment
LED INDICATION
NOTE The Battery status is determined by the LEDs in the status window.
Green Flashing
Green Solid
Blue Flashing
Blue Solid
Red Flashing
Red Solid
BATTERY RECALIBRATION INSTRUCTIONS
NOTES
During recalibration the Charger may become warm.
The calibration cycle will perform best if the Battery has a full charge in the beginning.
The Battery can either be calibrated and charged or just charged. Calibration takes longer than just charging.
It is recommended that the recalibration button is not pressed after the calibration has started.
1. When calibration is required, the red LED will flash when the Battery is placed into the Charger bay.
NOTE Recalibration is used to reset the Battery gauge to full accuracy.
2. Press the recalibration button on the front of the Charger (see Figure 9) to recalibrate the Battery. If the button is not pressed the Charger
will automatically charge the Battery.
3. When the blue LED flashes, the Battery is recalibrating.
4. When the calibration is completed, a solid blue LED will be present.
TotalShield II
Battery Charger
Battery Charging
Battery Fully Charged
Battery in calibration mode
Battery fuel gauge calibrated